Preliminary design of a tangentially viewing imaging bolometer for NSTX-U

  • B. J. Peterson
  • , R. Sano
  • , M. L. Reinke
  • , J. M. Canik
  • , L. F. Delgado-Aparicio
  • , J. D. Lore
  • , K. Mukai
  • , T. K. Gray
  • , G. G. Van Eden
  • , M. A. Jaworski

Research output: Contribution to journalArticlepeer-review

4 Scopus citations

Abstract

The infrared imaging video bolometer (IRVB) measures plasma radiated power images using a thin metal foil. Two different designs with a tangential view of NSTX-U are made assuming a 640 × 480 (1280 × 1024) pixel, 30 (105) fps, 50 (20) mK, IR camera imaging the 9 cm × 9 cm × 2 μm Pt foil. The foil is divided into 40 × 40 (64 × 64) IRVB channels. This gives a spatial resolution of 3.4 (2.2) cm on the machine mid-plane. The noise equivalent power density of the IRVB is given as 113 (46) μW/cm2 for a time resolution of 33 (20) ms. Synthetic images derived from Scrape Off Layer Plasma Simulation data using the IRVB geometry show peak signal levels ranging from ∼0.8 to ∼80 (∼0.36 to ∼26) mW/cm2.

Original languageEnglish (US)
Article number11D140
JournalReview of Scientific Instruments
Volume87
Issue number11
DOIs
StatePublished - Nov 1 2016

All Science Journal Classification (ASJC) codes

  • Instrumentation

Fingerprint

Dive into the research topics of 'Preliminary design of a tangentially viewing imaging bolometer for NSTX-U'. Together they form a unique fingerprint.

Cite this