Methods and preliminary measurement results of liquid Li wettability

G. Z. Zuo, J. S. Hu, J. Ren, Z. Sun, Q. X. Yang, J. G. Li, L. E. Zakharov, D. K. Mansfield

Research output: Contribution to journalReview articlepeer-review

20 Scopus citations

Abstract

A test of lithium wettability was performed in high vacuum (< 3 × 10-4 Pa). High magnification images of Li droplets on stainless steel substrates were produced and processed using the MATLAB® program to obtain clear image edge points. In contrast to the more standard "θ/2" or polynomial fitting methods, ellipse fitting of the complete Li droplet shape resulted in reliable contact angle measurements over a wide range of contact angles. Using the ellipse fitting method, it was observed that the contact angle of a liquid Li droplet on a stainless steel substrate gradually decreased with increasing substrate temperature. The critical wetting temperature of liquid Li on stainless steel was observed to be about 290 C.

Original languageEnglish (US)
Article number023506
JournalReview of Scientific Instruments
Volume85
Issue number2
DOIs
StatePublished - Feb 2014
Externally publishedYes

All Science Journal Classification (ASJC) codes

  • Instrumentation

Fingerprint

Dive into the research topics of 'Methods and preliminary measurement results of liquid Li wettability'. Together they form a unique fingerprint.

Cite this