Abstract
We have developed a low-cost, robust, multifoil-filtered spectrometer to provide absolute measurements of low-Z impurity concentrations in the Madison Symmetric Torus reversed-field pinch. The spectrometer utilizes an array of six thin-film coated soft x-ray diodes. Each multilayered coating is specifically tailored to isolate the K-shell emission lines of H- and He-like oxygen, carbon, and aluminum. With calibrations obtained via a synchrotron source, absolute measurements of photon flux have been made. We address the technical aspects of this diagnostic and present impurity data from both standard and high-confinement plasma discharges.
| Original language | English (US) |
|---|---|
| Pages (from-to) | 1188-1191 |
| Number of pages | 4 |
| Journal | Review of Scientific Instruments |
| Volume | 72 |
| Issue number | 1 II |
| DOIs | |
| State | Published - Jan 2001 |
| Externally published | Yes |
All Science Journal Classification (ASJC) codes
- Instrumentation