The charge exchange recombination spectroscopy diagnostic on the DIII-D tokamak has been upgraded with the addition of more high radial resolution view chords near the edge of the plasma (r/a > 0.8). The additional views are diagnosed with the same number of spectrometers by placing fiber optics side-by-side at the spectrometer entrance with a precise separation that avoids wavelength shifted crosstalk without the use of bandpass filters. The new views improve measurement of edge impurity parameters in steep gradient, H-mode plasmas with many different shapes. The number of edge view chords with 8 mm radial separation has increased from 16 to 38. New fused silica fibers have improved light throughput and clarify the observation of non-Gaussian spectra that suggest the ion distribution function can be non-Maxwellian in low collisionality plasmas.
| Original language | English (US) |
|---|---|
| Article number | 11E512 |
| Journal | Review of Scientific Instruments |
| Volume | 87 |
| Issue number | 11 |
| DOIs | |
| State | Published - Nov 1 2016 |
All Science Journal Classification (ASJC) codes
- Instrumentation
Fingerprint
Dive into the research topics of 'Improved edge charge exchange recombination spectroscopy in DIII-D'. Together they form a unique fingerprint.Cite this