Diagnostic performance of the beam emission spectroscopy system on the National Spherical Torus Experiment

Research output: Contribution to journalArticlepeer-review

Abstract

The beam emission spectroscopy system on the National Spherical Torus Experiment measures localized density fluctuations on the ion gyroscale. Optical sightlines provide core to edge radial coverage, and the sightlines are aligned to typical pitch angles to maximize cross-field spatial resolution. Sightline images are 2-3 cm, and point spread function calculations indicate image distortion from pitch angle misalignment and atomic state finite lifetimes is minor with a 15% increase in the image size. New generation photodetectors achieve photon noise limited measurements at frequencies up to 400 kHz with refrigerant cooling at -20 °C. Measurements near the pedestal show broadband turbulence up to 100 kHz, and poloidal correlation lengths are about 10 cm. Plasma turbulence signals can be 2-3 orders of magnitude above photon noise and amplifier thermal noise.

Original languageEnglish (US)
Article number10D502
JournalReview of Scientific Instruments
Volume83
Issue number10
DOIs
StatePublished - Oct 2012
Externally publishedYes

All Science Journal Classification (ASJC) codes

  • Instrumentation

Fingerprint

Dive into the research topics of 'Diagnostic performance of the beam emission spectroscopy system on the National Spherical Torus Experiment'. Together they form a unique fingerprint.

Cite this