Abstract
This article describes the overall design of the Joint European Torus (JET) motional Stark effect (MSE) diagnostic. The principle of this diagnostic has been proven on other tokamaks (TFTR and DIII-D) but its implementation on JET involves specific difficulties associated with optical access and neutral beam geometry. The solutions adopted for the access problem include the use of a novel "Air Amici" prism for beam deviation and low Verdet coefficient glass for the relay optics. MSE signals from the different beam sources are distinguished by their Doppler shift; the diagnostic uses the π+ components of the Stark spectrum to avoid contamination with signals from different beam sources. The optimum viewing location for good space resolution across the plasma profile is also one that allows good accuracy in the reconstruction of the q profile from the measurements. An assessment of the optical throughput predicts a measurement accuracy of the polarization angle of better than 0.15° with 1-10 ms time resolution.
| Original language | English (US) |
|---|---|
| Pages (from-to) | 894-897 |
| Number of pages | 4 |
| Journal | Review of Scientific Instruments |
| Volume | 70 |
| Issue number | 1 II |
| DOIs | |
| State | Published - Jan 1999 |
All Science Journal Classification (ASJC) codes
- Instrumentation
Fingerprint
Dive into the research topics of 'Design of the Joint European Torus motional stark effect diagnostic'. Together they form a unique fingerprint.Cite this