Channel electron multiplier compatibility with Viton and Apiezon-L vacuum grease

D. J. McComas, J. R. Baldonado, S. J. Bame, B. L. Barraclough

Research output: Contribution to journalArticlepeer-review

7 Scopus citations

Abstract

Clean Viton and Viton coated with Apiezon-L vacuum grease were tested for their noncontaminating compatibility with channel electron multipliers (CEMs). The test setup and procedure were the same as those used previously in conjunction with CEM compatibility tests of certain epoxies, solder, and fluorocarbon polymer materials useful for construction of spaceflight sensors. While some CEM gain degradation was noted during exposure to Viton and Apiezon-L, the present tests indicate that, at least over instrument lifetimes of ∼2×1012 counts, these materials should be suitable for (1) preflight space sensor testing systems, (2) hermetic seals for CEM-based space sensors, and (3) terrestrial CEM-based instrumentation.

Original languageEnglish (US)
Pages (from-to)2331-2332
Number of pages2
JournalReview of Scientific Instruments
Volume58
Issue number12
DOIs
StatePublished - 1987
Externally publishedYes

All Science Journal Classification (ASJC) codes

  • Instrumentation

Fingerprint

Dive into the research topics of 'Channel electron multiplier compatibility with Viton and Apiezon-L vacuum grease'. Together they form a unique fingerprint.