Carbon dioxide and ammonia detection using 2 μm diode laser based quartz-enhanced photoacoustic spectroscopy

R. Lewicki, G. Wysocki, A. A. Kosterev, F. K. Tittel

Research output: Contribution to journalArticlepeer-review

118 Scopus citations

Abstract

Quartz-enhanced photoacoustic spectroscopy was employed for trace gas concentration measurements of CO2 and NH3 using a continuous wave thermoelectrically cooled, distributed feedback diode laser operating at 2 μm. A normalized noise equivalent absorption coefficient, NNEA(1σ)=1.4×10-8 cm-1W/ √Hz was obtained for CO2 using the R18 line of the 2ν13 band at 4991.26 cm-1. This corresponds to minimum detection limit (1σ) of 18 parts per million (ppm) for a 1 s lock-in time constant. The influence of the H2O presence in the sample gas mixture on the CO2 sensor performance was investigated. Ammonia detection was performed using the P P 6(6)S line of the ν34 band at 4986.99 cm-1. A detection limit (1σ) of 3 ppm for NH3 concentration with a 1 s lock-in time constant was achieved. This results in a normalized noise equivalent absorption of NNEA(1σ)=8.9×10-9 cm-1W/ √Hz.

Original languageEnglish (US)
Pages (from-to)157-162
Number of pages6
JournalApplied Physics B: Lasers and Optics
Volume87
Issue number1
DOIs
StatePublished - Mar 2007