Abstract
We have developed a fast-scanning heterodyne receiver into a plasma diagnostic that measures the fundamental cyclotron emission from the PLT plasma and thus ascertains the time evolution of the electron temperature profile. The receiver scans 60-90 GHz every 10 ms and is interfaced to a computer for completely automated calibrated temperature measurements.
| Original language | English (US) |
|---|---|
| Pages (from-to) | 949-951 |
| Number of pages | 3 |
| Journal | Review of Scientific Instruments |
| Volume | 50 |
| Issue number | 8 |
| DOIs | |
| State | Published - 1979 |
All Science Journal Classification (ASJC) codes
- Instrumentation